Model 613D: Has a 30 minute mechanical timer, protective circuit breaker and a lid safety
switch that cuts power to the motor if the lid latch is not secure. The 613D also
features a higher RPM motor resulting in increased speed and force.
It is recommended that your Model 613 centrifuge be calibrated at least every 6 months.
The max speed for the 613E should fall between 3000 and 3200 RPM, this will give a
separation force of 1090 - 1240 xg. The max speed of the 613D should fall between 3250
and 3400, this will give a separation force of 1275 - 1400 xg.
A phototachometer is suggested to determine RPM.
Each model 613 centrifuge comes standard with a 6-place, 45-degree angled rotor and
six (6) test tube holders for use with both 15mL and 10mL test tubes. Six (6) test tube
cushions, (part no. 1525), are included for spinning 10mL test tubes (12mm x 100mm);
15mL test tubes, (16mm x 125mm), do not require test tube cushions. Optional test tube
holders are available for 10mL (100mm) and 5mL (75mm) test tubes that require no tube
Environmental Conditions: This centrifuge is not intended for outside use or for use in
extreme environmental conditions. Any use other than the manufacturer s suggested
usage may impair the protection provided by the unit.